KE

KEYENCE CORPORATION NR-HV04

NR-500 series (Multi-input Data Logger) - NR-HV04

  • Бренд: KEYENCE CORPORATION
  • Артикул (MPN): NR-HV04
  • Серия: NR-X Series
  • Статус производства: В производстве

Категории:

Технические характеристики

TypeHigh-speed, high-voltage measurement unit
Input methodInsulated single-end input, channels insulated from each other, input channels insulated from other units (PC)
Number of channels4 ch
Measurement cycleMaximum sampling cycle: 1 μs (1 MHz) to 60 s
A/D converting systemSuccessive approximation, simultaneous sampling of all channels
A/D resolution14 bit
Input frequency band250 kHz (−3 dB or more)
Input type±1000 V, ±500 V, ±200 V, ±100 V, ±50 V, ±20 V, ±10 V, ±5 V, ±2 V
Low pass filterPASS, 7 Hz, 500 Hz, 50 kHz (−12 dB/oct)
Напряжение питания[When using OP-89263 1:1 voltage-resistant probe] 1000 V peak and 700 Vrms, or 700 VDC; [When using OP-89264 1:1 voltage-resistant probe] 300 VDC + AC peak; [When using OP-88998 10:1 voltage-resistant probe] 1000 VDC + AC peak; [Direct input] 60 VDC + AC peak
Digital filterAveraging rate: 1 to 128 (automatically selected)
Напряжение питания2200 VAC (50/60 Hz) for 1 minute
Connector formatInsulated BNC connector
Input impedance1 MΩ ±1%, approx. 30 pF
Input couplingAC/DC/AC-RMS/DC-RMS
Measurable range±1000 V: −1000.00 to +1000.00 V ±500 V: −550.00 to +550.00 V ±200 V: −220.00 to +220.00 V ±100 V: −110.000 to +110.000 V ±50 V: −55.000 to +55.000 V ±20 V: −22.000 to +22.000 V ±10 V: −11.0000 to +11.0000 V ±5 V: −5.5000 to +5.5000 V ±2 V: −2.2000 to +2.2000 V
Measurement accuracy±0.1% of F.S., LPF: 500 Hz, with digital filter ON
Display resolution±1000 V: 50 mV ±500 V: 20 mV ±200 V: 10 mV ±100 V: 5 mV ±50 V: 2 mV ±20 V: 1 mV ±10 V: 0.5 mV ±5 V: 0.2 mV ±2 V: 0.1 mV
Buffer memory16M data
Warm up time30 minutes or more
RMS measurement function (Number of channels)4 ch
Напряжение питания[When using OP-89263 1:1 voltage-resistant probe] Measurement terminal, GND terminal: 300 Vrms (CAT II); [When using OP-89264 1:1 voltage-resistant probe] Measurement terminal, GND terminal: 300 Vrms (CAT II); [When using OP-88998 10:1 voltage-resistant probe] Measurement terminal: 700 Vrms (CAT II), GND terminal: 300 Vrms (CAT II); [Direct input] Measurement terminal, GND terminal: 42.4 Vpeak and 30 Vrms, or 60 VDC
RMS measurement function (Accuracy)Sine wave: 20 Hz to 1 kHz, ±0.5% of F.S.; 1 to 20 kHz, ±1.5% of F.S.; Crest factor: 4 or less; Response time: 500 ms
Residual noise level0.01% of F.S., LPF: 500 Hz (200 μs/S, digital filter: ON) (typ.); 0.05% of F.S., LPF: PASS (typ.)
CalculationfunctionCH1 to CH2, CH3 to CH4 (Single-channel usage not possible when using the calculation function)
Common mode noise reduction ratio80 dB or more (at max. 60 Hz DC) (typ.)
Rating (Power consumption)3 W or less
Environmental resistance (Insulation resistance)500 VDC, 10 MΩ or more
Рабочая температура0 to +40°C 32°F to 104°F
Environmental resistance (Relative humidity)10 to 85% RH (no condensation)
МассаApprox. 270 g 9.53 oz
Запросить КП по этой позиции