| Type | High-speed, high-voltage measurement unit |
|---|---|
| Input method | Insulated single-end input, channels insulated from each other, input channels insulated from other units (PC) |
| Number of channels | 4 ch |
| Measurement cycle | Maximum sampling cycle: 1 μs (1 MHz) to 60 s |
| A/D converting system | Successive approximation, simultaneous sampling of all channels |
| A/D resolution | 14 bit |
| Input frequency band | 250 kHz (−3 dB or more) |
| Input type | ±1000 V, ±500 V, ±200 V, ±100 V, ±50 V, ±20 V, ±10 V, ±5 V, ±2 V |
| Low pass filter | PASS, 7 Hz, 500 Hz, 50 kHz (−12 dB/oct) |
| Напряжение питания | [When using OP-89263 1:1 voltage-resistant probe] 1000 V peak and 700 Vrms, or 700 VDC; [When using OP-89264 1:1 voltage-resistant probe] 300 VDC + AC peak; [When using OP-88998 10:1 voltage-resistant probe] 1000 VDC + AC peak; [Direct input] 60 VDC + AC peak |
| Digital filter | Averaging rate: 1 to 128 (automatically selected) |
| Напряжение питания | 2200 VAC (50/60 Hz) for 1 minute |
| Connector format | Insulated BNC connector |
| Input impedance | 1 MΩ ±1%, approx. 30 pF |
| Input coupling | AC/DC/AC-RMS/DC-RMS |
| Measurable range | ±1000 V: −1000.00 to +1000.00 V ±500 V: −550.00 to +550.00 V ±200 V: −220.00 to +220.00 V ±100 V: −110.000 to +110.000 V ±50 V: −55.000 to +55.000 V ±20 V: −22.000 to +22.000 V ±10 V: −11.0000 to +11.0000 V ±5 V: −5.5000 to +5.5000 V ±2 V: −2.2000 to +2.2000 V |
| Measurement accuracy | ±0.1% of F.S., LPF: 500 Hz, with digital filter ON |
| Display resolution | ±1000 V: 50 mV ±500 V: 20 mV ±200 V: 10 mV ±100 V: 5 mV ±50 V: 2 mV ±20 V: 1 mV ±10 V: 0.5 mV ±5 V: 0.2 mV ±2 V: 0.1 mV |
| Buffer memory | 16M data |
| Warm up time | 30 minutes or more |
| RMS measurement function (Number of channels) | 4 ch |
| Напряжение питания | [When using OP-89263 1:1 voltage-resistant probe] Measurement terminal, GND terminal: 300 Vrms (CAT II); [When using OP-89264 1:1 voltage-resistant probe] Measurement terminal, GND terminal: 300 Vrms (CAT II); [When using OP-88998 10:1 voltage-resistant probe] Measurement terminal: 700 Vrms (CAT II), GND terminal: 300 Vrms (CAT II); [Direct input] Measurement terminal, GND terminal: 42.4 Vpeak and 30 Vrms, or 60 VDC |
| RMS measurement function (Accuracy) | Sine wave: 20 Hz to 1 kHz, ±0.5% of F.S.; 1 to 20 kHz, ±1.5% of F.S.; Crest factor: 4 or less; Response time: 500 ms |
| Residual noise level | 0.01% of F.S., LPF: 500 Hz (200 μs/S, digital filter: ON) (typ.); 0.05% of F.S., LPF: PASS (typ.) |
| Calculationfunction | CH1 to CH2, CH3 to CH4 (Single-channel usage not possible when using the calculation function) |
| Common mode noise reduction ratio | 80 dB or more (at max. 60 Hz DC) (typ.) |
| Rating (Power consumption) | 3 W or less |
| Environmental resistance (Insulation resistance) | 500 VDC, 10 MΩ or more |
| Рабочая температура | 0 to +40°C 32°F to 104°F |
| Environmental resistance (Relative humidity) | 10 to 85% RH (no condensation) |
| Масса | Approx. 270 g 9.53 oz |