KE

KEYENCE CORPORATION NR-ST04

NR-500 series (Multi-input Data Logger) - NR-ST04

  • Бренд: KEYENCE CORPORATION
  • Артикул (MPN): NR-ST04
  • Серия: NR-X Series
  • Статус производства: В производстве

Категории:

Технические характеристики

TypeStrain measurement unit
Input methodBalanced differential input, channels not insulated from each other, input channels insulated from other units (PC)
Number of channels4 ch
Measurement cycleMaximum sampling cycle: 20 μs (50 kHz) to 60 s
A/D converting systemSuccessive approximation, simultaneous sampling of all channels
A/D resolution16 bit
Input frequency band5 kHz DC max. (−3 dB)
Maximum input rating±1.0 V
Напряжение питания1500 VAC (50/60 Hz) for 1 minute
Input impedance1 MΩ or more
Напряжение питания±2 mV, ±5 mV, ±10 mV, ±20 mV, ±50 mV
Input type (Strain input)±1000 μST, ±2000 μST, ±5000 μST, ±10000 μST, ±20000 μST
Measurable range±50 mV: −55.00 to +55.00 mV ±20 mV: −22.000 to +22.000 mV ±10 mV: −11.000 to +11.000 mV ±5 mV: −5.500 to +5.500 mV ±2 mV: −2.2000 to +2.2000 mV ±20000 μST: −22000 to +22000 μST ±10000 μST: −11000.0 to +11000.0 μST ±5000 μS: −5500.00 to +5500.00 μST ±2000 μST: −2200.0 to +2200.0 μST ±1000 μST: −1100.00 to +1100.00 μST
Measurement accuracy (Gain accuracy)±0.2% of F.S. (with low-pass filter of 8 Hz and 50 Hz sampling averaged 20 times)
Display resolution±50 mV: 10 μV ±20 mV: 1 μV ±10 mV: 1 μV ±5 mV: 1 μV ±2 mV: 0.1 μV ±20000 μST: 1 μST ±10000 μST: 0.5 μST ±5000 μS: 0.25 μST ±2000 μST: 0.1 μST ±1000 μST: 0.05 μST
Buffer memory4M data
Warm up time30 minutes or more
Applicable gauge resistanceQuarter bridge / half bridge system: 120 Ω (A 350 Ω strain gauge requires an external bridge box.); Full bridge system: 120 Ω to 1 kΩ
Напряжение питания2 VDC (±0.4%)
Balance adjustment (System)Electronic balancing (auto-balancing)
Balance adjustment (Balance adjustment accuracy)±0.1% of F.S. (with low-pass filter of 8 Hz and 50 Hz sampling averaged 20 times)
Balance adjustment (Balance adjustment range)±29000 μST (during strain input), ±29 mV (during voltage input)
Gauge ratio2.0 (fixed)
Low pass filter (Cutoff frequency)Auto, 5 kHz, 3.5 kHz (3 kHz), 2 kHz (1.5 kHz), 1 kHz (750 Hz), 500 Hz (370 Hz), 250 Hz (200 Hz), 120 Hz (100 Hz), 60 Hz (50 Hz), 30 Hz (25 Hz), 15 Hz (12 Hz), 8 Hz (6 Hz), 4 Hz (3 Hz); Value in parentheses is for 20 kHz sampling.
Low pass filter (Attenuation characteristics)5th order Bessel filter, −30 dB/oct
Zero-point stability (Temperature characteristics)±2 μST/°C, ±2 μV/°C
Zero-point stability (Chronological changes)±1.5 μST/8 h, ±3 μV/8 h
Gain stability (Temperature characteristics)±0.05% of F.S./°C
Gain stability (Chronological changes)±0.3% of F.S./8 h
Common mode noise reduction ratio80 dB or more (60 Hz DC max.)
Rating (Power consumption)3.0 W or less
Рабочая температура0 to +40°C 32°F to 104°F
Environmental resistance (Relative humidity)10 to 85% RH (no condensation)
МассаApprox. 230 g 8.12 oz (including two OP-51584: NR-ST04 10-pole terminal blocks)
Запросить КП по этой позиции