KE

KEYENCE CORPORATION NR-TH08

NR-500 series (Multi-input Data Logger) - NR-TH08

  • Бренд: KEYENCE CORPORATION
  • Артикул (MPN): NR-TH08
  • Серия: NR-X Series
  • Статус производства: В производстве

Категории:

Технические характеристики

TypeHigh-accuracy temperature/voltage measurement unit
Input methodFloating unbalanced input, channels insulated from each other, CH inputs insulated from other units (PC), channels not insulated from each other for resistance thermometers
Number of channelsInput: 8
Measurement cycleFastest sampling cycle: 100 ms (A/D integration time: 2 ms) to 1 h
A/D converting systemDelta-sigma method
A/D resolution24 bit
A/D integration time2 ms, 16.7 ms, 20 ms
Input typeVoltage: ±50 V, ±10 V, ±5 V, ±1 V, ±0.5 V, ±0.1 V; Thermocouple; K, J, E, T, R, S, B, N, C; Resistance thermometer: Pt100, JPt100 3-line type; Specified current: 1 mA
Measurement range (±50 V)Measurable range: −55.00 to +55.00 V Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±2 digits Display resolution: 10 mV
Measurement range (±10 V)Measurable range: −11.000 to +11.000 V Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±2 digits Display resolution: 1 mV
Measurement range (±5 V)Measurable range: −5.500 to +5.500 V Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±2 digits Display resolution: 1 mV
Measurement range (±1 V)Measurable range: −1.1000 to +1.1000 V Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±2 digits Display resolution: 0.1 mV
Measurement range (±500 mV)Measurable range: −550.0 to +550.0 mV Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±2 digits Display resolution: 0.1 mV
Measurement range (±100 mV)Measurable range: −110.00 to +110.00 mV Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±2 digits Display resolution: 0.01 mV
Measurement range (K)Measurable range: −100 to +1372°C -148°F to +2501.6°F , −200 to −100°C -328°F to -148°F Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.6°C ±1.1°F , ±0.05% of rdg ±0.9°C ±1.6°F Display resolution: 0.05°C 0.09°F
Measurement range (J)Measurable range: −100 to +1200°C -148°F to +2192°F , −200 to −100°C -328°F to -148°F Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.6°C ±1.1°F , ±0.05% of rdg ±0.8°C ±1.4°F Display resolution: 0.05°C 0.09°F
Measurement range (E)Measurable range: −100 to +1000°C -148°F to +1832°F , −200 to −100°C -328°F to -148°F Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.6°C ±1.1°F , ±0.05% of rdg ±0.8°C ±1.4°F Display resolution: 0.05°C 0.09°F
Measurement range (T)Measurable range: −100 to +400°C -148°F to +752°F , −200 to −100°C -328°F to -148°F Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.5°C ±0.9°F , ±0.05% of rdg ±0.8°C ±1.4°F Display resolution: 0.01°C 0.02°F
Measurement range (N)Measurable range: 0 to 1300°C 32°F to 2372°F Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.6°C ±1.1°F Display resolution: 0.05°C 0.09°F
Measurement range (C)Measurable range: 1500 to 2315°C 2732°F to 4199°F , 0 to 1500°C 32°F to 2732°F Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±1.1°C ±2.0°F , ±0.05% of rdg ±0.8°C ±1.4°F Display resolution: 0.05°C 0.09°F
Measurement range (R)Measurable range: 300 to 1768°C 572°F to 3214.4°F , 0 to 300°C 32°F to 572°F Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.8°C ±1.4°F , ±0.05% of rdg ±1.6°C ±2.9°F Display resolution: 0.05°C 0.09°F
Measurement range (S)Measurable range: 300 to 1768°C 572°F to 3214.4°F , 0 to 300°C 32°F to 572°F Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±0.9°C ±1.6°F , ±0.05% of rdg ±1.6°C ±2.9°F Display resolution: 0.05°C 0.09°F
Measurement range (B)Measurable range: 400 to 600°C 752°F to 1112°F , 600 to 1820°C 1112°F to 3308°F Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.05% of rdg ±1.7°C ±3.1°F , ±0.05% of rdg ±1.0°C ±1.8°F Display resolution: 0.05°C 0.09°F
Measurement range (Pt100)Measurable range: −200 to +660°C -328°F to +1220°F Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.1% of rdg ±0.3°C ±0.5°F Display resolution: 0.02°C 0.04°F
Measurement range (JPt100)Measurable range: −200 to +510°C -328°F to +950°F Measurement accuracy (A/D integration time: 16.7 ms, 20 ms): ±0.1% of rdg ±0.3°C ±0.5°F Display resolution: 0.02°C 0.04°F
Reference contact compensationSwitchable between internal and external (individual unit setting, 0°C 32°F for external)
Reference contact compensation accuracy±0.7°C ±1.3°F (23°C ±2°C 73.4°F ±3.6°F , at input terminal temperature equilibrium)
Напряжение питания±60 V (rated)
Напряжение питания10 MΩ or more
Напряжение питанияApprox. 1 MΩ
Напряжение питания1500 VAC (50/60 Hz) for 1 minute
Напряжение питания120 Vp-p AC/DC
Напряжение питания2 kΩ or less
Input signal source resistance (Resistance thermometer)10 Ω or less per line (three lines must be equal)
Thermocouple burnoutWire breakage detected in the thermocouple range by applying 1 μA of current
Burnout detection frequencyDetected in the measurement cycle
Buffer memory200k data
Warm up time30 minutes or more
Rating (Power consumption)1.2 W or less
Рабочая температура0 to +40°C 32°F to 104°F
Environmental resistance (Relative humidity)10 to 85% RH (no condensation)
МассаApprox. 230 g 8.12 oz (including approx. 100 g 3.53 oz for the OP-51582 (NR-TH08 terminal block))
Запросить КП по этой позиции