Технические характеристики

Input methodInsulated single-end input, channels insulated from each other, input channels insulated from the case and other units (system)
Number of input channels8
Input connector typeDedicated 3-pin connector
Measurement cycleMaximum sampling cycle: 100 μs (10 kHz) to 60 s
A/D converting systemDelta-sigma method
A/D resolution24 bits
A/D integration time (noise filter)AUTO (Automatic configuration according to sampling cycle), 25 μs (40 kHz), 208 μs (4.8 kHz), 2.5 ms (400 Hz), 16.7 ms (60 Hz), 20 ms (50 Hz), 100 ms (10 Hz)
Input frequency band2 kHz (−3 dB or more)
Input typeVoltage input: ±60 V, ±20 V, ±10 V, ±5 V, ±2 V, ±1 V, ±500 mV, ±200 mV, ±100 mV Current input: ±20 mA
AveragingSwitchable between ON/OFF, Averaging rate: 2 to 128 (moving average)
Напряжение питания1 MΩ ±2%
Напряжение питания10 MΩ ±2%
Input impedance (Current input)Approx. 100 Ω
Measurement range (±60 V)Measurable range: −60.000 to +60.000 V Measurement accuracy: ±0.05% of rdg ±8 mV Display resolution: 2 mV
Measurement range (±20 V)Measurable range: −22.000 to +22.000 V Measurement accuracy: ±0.05% of rdg ±4 mV Display resolution: 1 mV
Measurement range (±10 V)Measurable range: −11.000 to +11.000 V Measurement accuracy: ±0.05% of rdg ±2 mV Display resolution: 0.5 mV
Measurement range (±5 V)Measurable range: −5.5000 to +5.5000 V Measurement accuracy: ±0.05% of rdg ±0.8 mV Display resolution: 0.2 mV
Measurement range (±2 V)Measurable range: −2.2000 to +2.2000 V Measurement accuracy: ±0.05% of rdg ±0.4 mV Display resolution: 0.1 mV
Measurement range (±1 V)Measurable range: −1.1000 to +1.1000 V Measurement accuracy: ±0.05% of rdg ±0.2 mV Display resolution: 0.05 mV
Measurement range (±500 mV)Measurable range: −550.00 to +550.00 mV Measurement accuracy: ±0.05% of rdg ±0.08 mV Display resolution: 0.02 mV
Measurement range (±200 mV)Measurable range: −220.00 to +220.00 mV Measurement accuracy: ±0.05% of rdg ±0.04 mV Display resolution: 0.01 mV
Measurement range (±100 mV)Measurable range: −110.000 to +110.000 mV Measurement accuracy: ±0.05% of rdg ±0.02 mV Display resolution: 0.005 mV
Measurement range (±20 mA)Measurable range: −22.000 to +22.000 mA Measurement accuracy: ±0.08% of rdg ±0.004 mA Display resolution: 0.001 mA
Maximum input ratingVoltage input: ±60 V, Current input: ±30 mA
Напряжение питания120 VAC and 170 Vpeak, or 120 VDC (when using NR-XHA08T protective grounding) The following applies when not using NR-XHA08T protective grounding. Dry locations: 30 VAC and 42.4 Vpeak, or 60 VDC Wet locations: 16 VAC and 22.6 Vpeak, or 35 VDC
Напряжение питанияBetween measurement input terminal and case/other units (system): 1000 VAC (50 Hz / 60 Hz) for 1 minute Between measurement input terminals: 1000 VAC (50 Hz / 60 Hz) for 1 minute
Buffer memory16M data
Warm up time30 minutes or more
Residual noise level0.005% of F.S. (typ.) when A/D integration time is set to 20 ms (50 Hz) / 16.7 ms (60 Hz) 0.05% of F.S. (typ.) when A/D integration time is set to 208 μs (4.8 kHz)
Common mode noise reduction ratio105 dB or more (up to 60 Hz DC) (typ.)
Remote connectionDedicated environment-resistant remote unit input port × 1 (Remote IN), Output port × 1 (Remote OUT) Connection to NR-X100W master data collection unit or other NR-XR01 remote units using environment-resistant unit cable Connectable to multiple environment-resistant measurement units
Synchronization accuracy between remote connection units±1 μs
Power supplySupplied from NR-X100W master unit
Power consumption5.8 W or less
Surrounding air temperature−40 to +85°C -40°F to +185°F
Operating ambient humidityMax. 100% RH
Vibration resistance100 m/s 2 328.1 ft/s 2 ; 5 to 2000 Hz; 2 hours in each of the X, Y, and Z directions (IEC 60068-2-6) Equivalent to JIS D1601 Stage 45 (Class 1, Class 2, and Class 3; equivalent to Class B) automotive vibration standard
Shock resistance1000 m/s 2 3280.8 ft/s 2 ; 3 times in each of the X, Y, and Z directions (18 times in total) (IEC 60068-2-27)
Степень защиты (IP)IP65/IP67 (IEC 60529) Enclosure Type 4X (UL 50E, CSA C22.2 No. 94.2, NEMA 250) Indoor/outdoor use
МатериалMain unit case: Aluminum die-casting (coated), Display: Reinforced glass, Connectors: Brass (chromate processing), Connector protection cap: PBT+PC
МассаApprox. 690 g 24.36 oz
Запросить КП по этой позиции