GHP-456-TITAN-®

NETZSCH Group (Erich NETZSCH B.V. & Co. Holding KG) -- Pumps & Systems: NETZSCH Pumpen & Systeme GmbH; Grinding & Dispersing: NETZSCH-Feinmahltechnik GmbH; Analyzing & Testing: NETZSCH-Gerätebau GmbH GHP-456-TITAN-®

GHP 456 Titan ®

Технические характеристики

Technique/DesignAbsolute method (no calibration or reference materials required) Symmetrical arrangement Fully automated operation
StandardsBased on standards such as ISO 8302, ASTM C177, DIN/EN 12667, DIN/EN 12939, etc
Mean specimen temperature rangeLow‐temperature version: ‐160°C to 250°C High‐temperature version: ‐160°C to 600°C Both versions require LN2 cooling for the sub‐ambient temperature range
Cooling systemsLiquid nitrogen (LN 2 ): ‐160°C to 250°C Compressed air: 50°C to 300°C Chiller: 20°C to 85°C → No active cooling from 300°C to 600°C
Plate dimensions300 mm x 300 mm Motorized plate hoist
МатериалLow‐temperature version: Aluminum alloy High‐temperature version: Tungsten alloy
Plate temperature rangeStandard version: ‐180°C to 270°C High‐temperature version: ‐180°C to 620°C
Vacuum‐tightnessBy design, 5 x 10 ‐4 mbar (0.05Pa)
Defined pressure levelsControlled between 0.1 mbar and 100 mbar
Specimen thicknessUp to 100 mm (typically 10 ... 50 mm) Max. difference in thickness for the two specimens to be measured: 2%
Atmosphere/ pressure levelOxidizing up to 300°C Inert Vacuum Defined pressure levels
Thermal conductivity range0.003 to 2 W/(m·K)*
Minimum measurable thermal resistance0.02 m 2 ·K/W*
AccuracyTypically 2%*
ReproducibilityTypically < 1%*
Software specialtiesSmartMode , including: Method‐based, easy operation (e.g., user and predefined methods) Support of controlled, adaptive cooling Report generator Results including combined standard uncertainties according to GUM**
Получить предложение по этой позиции